Open-access Calibration and Maintenance of Low Power Laser Devices: Descriptive Study

Introduction:  The low-power laser is a technique used in several areas of health: physiotherapy, dentistry, otorhinolaryngology and dermatology, among others. To achieve therapeutic effects, the dosage is essential and depends on the energy levels emitted by the device.

Objective:  To know the status of calibration and maintenance of low power laser therapy devices, in physiotherapy services or counterparts of institutions in Montevideo - Uruguay.

Method:  A descriptive, cross-sectional study was conducted in physiotherapy services in Montevideo, Uruguay. 20 low-power laser devices were analyzed.

Results:  7 of the devices evaluated remained within the appropriate emission limits. Of the 13 lasers outside the limits to deliver adequate emission, 6 had a slight mismatch, 1 had a moderate mismatch, and 6 presented a severe mismatch. Among the total devices evaluated, only 6 had performed a previous calibration and only 2 of these were within the recommended times frames.

Conclusion:  A large number of low-power laser devices with imbalances in their emission was observed. In most cases, low-power laser devices are not properly maintained. Equipment should be timely calibrated in order to offer safe and effective therapy to users.

Keywords:
low-level light therapy; calibration; biomedical technology assessment

location_on
Facultad de Ciencias de la Salud - Universidad Católica del Uruguay. Av. 8 de Octubre 2738 Montevideo - Uruguay. CP 11600. Teléfono: (598) 2487 27 17 int. 6567 - E-mail: ecuidadoshumanizados@ucu.edu.uy
rss_feed Acompañe los números de esta revista en su lector de RSS
Ir para arriba